5-8
Infinity Vista
VF6
5 A
LARMS
AND
M
ESSAGES
Alarm Validation
The alarm validation feature minimizes nuisance alarms for transient conditions.
When a parameter value falls below or rises above its current alarm limits, the monitor
waits a predetermined time before triggering the alarm. If the parameter value returns
to within the upper or lower alarm limit before the end of the alarm validation period,
the monitor does not trigger the alarm.
The alarm validation period is part of the software (see table below) and cannot be
modified. The monitor triggers the alarm at the end of the validation period.
Parameters have the following alarm validation periods:
Parameter-Specific Alarm Validation Periods
Parameter
Upper Limit
(in seconds)
Lower Limit
(in seconds)
ECG/Heart rate (HR)
2
0
Pulse rate (PLS)
4
10
ST Segment Analysis*
60
60
Respiration Rate (Rsp)
8
10
Arterial Oxygen Saturation (SpO
2
)
4
10
T, PVC, Arrhythmia Events
0
0
Non-invasive blood pressure (NBP)
0
0
Invasive blood pressure
(ART, PA, CVP, ICP, GP1, GP2*)
4
4
*Available only if option is enabled
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...