B
IOMEDICAL
S
UPPORT
VF6
Infinity Vista
C-13
Diagnostic Logs
The diagnostic log records major changes that occur in the monitoring environment,
as well as operational errors.
The diagnostic log stores the most recent 200 errors or conditions, along with date and
time of the occurrence. The most recent messages are displayed first. Turn the rotary
knob to scroll through all entries. The codes that appear in the log help DrägerService
identify the cause of error or condition.
STEPS: Displaying the Diagnostic Log
1. Press the
Menu
fixed key.
2. Click on
Monitor Setup
.
3. Click on
Biomed
.
4. Enter the Biomed password and click the knob.
5. Click on
Diagnostic Logs
.
6. The monitor displays the diagnostic log. Turn the rotary knob clockwise to
scroll back to previous messages. Turn the knob counterclockwise to scroll
forward to current messages.
Changing Units of Measure
You can change the units of measure for the following parameters:
z
Temperature — °C (default) or °F
z
Pressures — mmHg (default) or kPa
z
ST — mm (default) or mV
Upon changing the unit of measure, the monitor displays the New Patient? prompt.
You must admit a new patient and thereby discharge the current patient, or the change
of units does not take effect.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...