B-2
Infinity Vista
VF6
B C
LEANING
AND
D
ISINFECTING
Cleaning and Disinfecting
Clean the monitor and all accessories after each patient or daily according to your
hospital’s standard procedures. We recommend the following cleaning solutions and
procedures.
Monitor
z
Clean the monitor with a gauze moistened in a soap solution.
z
Dry thoroughly with a lint-free cloth.
z
Disinfect the monitor with a gauze moistened with diluted alcohol.
z
Dry thoroughly with a lint-free cloth.
Patient Cables
z
Clean the patient cables with a gauze pad moistened with a soap solution.
z
Dry thoroughly with a lint-free cloth.
z
To disinfect patient cables, wipe the cables with a gauze moistened with
diluted alcohol.
z
Dry thoroughly with a lint-free cloth.
CAUTION:
Do not use disinfectants that contain phenol as they can spot plastics.
Do
not autoclave
or clean accessories with strong aromatic, chlorinated, ketone, ether, or
ester solvents. Never immerse electrical connectors.
CAUTION:
The material used for the monitor’s enclosure is a highly resistant
thermoplastic. Do not use plastic solvents, sharp tools or abrasives to clean it.
CAUTION:
Do not steam autoclave, gas sterilize, or immerse the monitor in water or
cleaning solutions. Do not subject the monitor to intense vacuum.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...