Saving a Patient Setup
You can save and re-use the current monitoring configuration. This saved
configuration is automatically restored when a new patient of the same category (e.g.,
pediatric) is admitted.
The monitor saves the following settings: Waveform channel assignments and scales,
alarm limits and on/off status, IBP pressure labels, arrhythmia monitoring settings
(adult or pediatric mode only), respiration mode and markers, apnea time, coincidence
alarm and pacemaker setting and detection.
STEPS: Saving a Patient’s Setup
1. Press the
Menu
fixed key.
2. Click on
Monitor Setup
.
3. Click on
Biomed
.
4. Dial in the Biomed password and click the knob.
5. Click on
Save Setups
.
6. Click on
Save Setup
, select
Confirm
(or
Cancel
) and click the knob.
The monitor saves the current monitoring setup for future use. The setup is available
even after the current patient has been discharged from this monitor.
Locked Options and Demo Mode
The Locked Options menu gives you access to the Demo mode and to a number of
monitoring options that are available for purchase. These options and their order
numbers are listed in the appendix
Options and Accessories
.
Using the Demo Mode
The monitor provides a demonstration mode to allow the presentation of features and
functions during a product introduction. Before demonstrating or testing the monitor,
disconnect all patient cables from the monitor.
NOTE:
You cannot save a setup when an OCRG is displayed on the screen. Exit the
OCRG display before saving.
NOTE:
In Demo mode, the All Alarms Off key is without function and creates an
error tone when pressed.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...