D
ATA
T
RANSFER
VF6
Infinity Vista
4-9
4. Click on
Care Unit
. Selection
***
is the default and automatically selects all
care units. If you want to narrow your search for the transfer bedside monitor
and select the specific care unit in which the
source
monitor is located, dial in
the appropriate care unit from the available choices in the menu and click the
knob.
5. Click on
Transfer Bed
.
6. Select the bed label of the
source
monitor and click the knob.
7. Click on
Start Transfer
.
8. Click on
Confirm
to start the transfer.
During transfer, the Transfer menu displays the status message
Transferring data.
Once the transfer is complete, the patient is automatically discharged from the source
monitor and admitted to the destination monitor.
If the transfer is interrupted or unsuccessful, the destination monitor displays a
transfer error message. In this case, the source monitor retains its data and you can
attempt the transfer again.
NOTE:
The menu shows only those care units which have at least one bedside
monitor in Standby mode.
NOTE:
z
The Transfer Bed menu displays all networked bedside monitors in the selected
care unit that are currently in Standby mode.
z
You must click on a bed label, even if there is only one available bedside
monitor listed in the transfer menu; if you do not select a bedside monitor, the
transfer cannot take place.
z
If a previous patient had not been discharged at the destination monitor before
data transfer, the discharge takes place automatically during transfer and all
previous patient data is deleted.
Содержание Infinity Vista
Страница 1: ...Instructions for Use Infinity Vista Infinity Patient Monitoring Series ...
Страница 2: ......
Страница 16: ...xiv Infinity Vista VF6 This page intentionally left blank ...
Страница 25: ...FRONT PANEL VF6 Infinity Vista 1 9 Front Panel 1 Fixed Keys 3 Power ON OFF Key 2 Rotary Knob ...
Страница 27: ...LEFT SIDE PANEL VF6 Infinity Vista 1 11 Left Side Panel 1 NBP Hose Connection 2 MULTIMED NEOMED Connection ...
Страница 58: ...2 16 Infinity Vista VF6 2 MONITOR SETUP This page intentionally left blank ...
Страница 66: ...3 8 Infinity Vista VF6 3 NETWORK APPLICATIONS This page intentionally left blank ...
Страница 80: ...4 14 Infinity Vista VF6 4 ADMISSION DISCHARGE TRANSFER This page intentionally left blank ...
Страница 106: ...5 26 Infinity Vista VF6 5 ALARMS AND MESSAGES This page intentionally left blank ...
Страница 107: ...6 Trends Overview 6 2 Trend Setup 6 2 Trend Graphs 6 3 Trend Table 6 5 Special Conditions and Codes 6 6 ...
Страница 114: ...6 8 Infinity Vista VF6 6 TRENDS This page intentionally left blank ...
Страница 166: ...10 8 Infinity Vista VF6 10 ST SEGMENT ANALYSIS This page intentionally left blank ...
Страница 204: ...13 12 Infinity Vista VF6 13 NON INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Страница 213: ...15 Temperature Overview 15 2 Temperature Probes 15 2 Placing the Probe 15 2 ...
Страница 216: ...15 4 Infinity Vista VF6 15 TEMPERATURE This page intentionally left blank ...
Страница 250: ...C 18 Infinity Vista VF6 C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT This page intentionally left blank ...
Страница 272: ...E 4 Infinity Vista VF6 E GLOSSARY This page intentionally left blank ...