CPI Canada Inc.
Troubleshooting
5
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-8
5.2.3
Error Messages (Cont)
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touchscreen console)
PROBLEM
ACTION
E043
KV ERROR
High voltage error - kV
detected in non X-ray
state
kV detected in non X-ray state.
Switch OFF generator.
Prevent further use of
generator. Call product
support.
E044
COMM ERROR
Invalid communication
message
Received communication message
not valid and ignored.
Reset error.
E045 NOT
SUPPORTED Communication
message not supported
Received message is valid, but not
supported by this system.
Reset error.
E046
MODE INHIBITED
Mode Inhibited
Received message is valid, but not
allowed during present state.
Reset error.
E049
NOT ENABLED
Not enabled
Requested function not programmed
to be enabled.
Reprogram to enable function.
E050
GEN DATA ERROR
Generator limit data
corrupt
Generator detects corrupt generator
limit data.
Reprogram generator limit
data or set factory defaults.
E051
AEC DEVICE ERR
AEC feedback error (no
feedback signal
detected)
Generator has detected no, or
insufficient, feedback signal from the
AEC device.
1. Check that X-ray tube is
pointing at correct AEC
device.
2. Check AEC cable for
damage and proper
connection.
E052
HIGH SF CURRENT
High small focus filament
current error in standby
Generator detects small focus
filament current greater than limits in
standby mode.
Check filament board.
E053
HIGH LF CURRENT
High large focus filament
current error in standby
Generator detects large focus
filament current greater than limits in
standby mode.
Check filament board.
E054
AEC OUT OF RANGE
AEC reference out of
range
AEC reference has reached a
maximum or minimum limit.
Re-adjust AEC calibration
including density to operate
within AEC range (O to 1O
VDC).