CPI Canada Inc.
Troubleshooting
5
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-7
5.2.3
Error Messages (Cont)
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touchscreen console)
PROBLEM
ACTION
E017
NOT CALIBRATED
Uncalibrated Exposure
Parameter
Selected mA not calibrated for
selected kV.
Recalibrate X-ray tube.
E018
PREP TIMEOUT
Preparation Timeout
Generator has been in prep state too
long.
Reduce length of time in prep
state.
E019
ANODE HEAT LIMIT
Anode Heat Limit
Selected parameters will cause X-
ray tube to exceed its programmed
anode heat limit.
Reduce parameters or wait for
tube to cool.
E020
THERMAL INT #1
Thermal Switch Interlock
#1 Error
X-ray tube # 1 too hot and its thermal
switch has opened.
Wait for X-ray tube # 1 to cool.
E022
DOOR INTERLOCK
Door Interlock Error
Door is open.
Close door.
E028
PREP SW CLOSED
Prep Input Active During
Initialization Phase
Prep input active during power on
initialization phase.
Check prep switch and input
for short circuit.
E029
X-RAY SW CLOSED
X-ray Input Active During
Initialization Phase
X-ray input active during power on
initialization phase.
Check X-ray switch and input
for short circuit.
E032 CONSOLE
COMM
ERR
Console Communication
Error
Generator has detected error in
communication to console.
1. Check console cable for
damage and proper
connection.
2. Turn power off and then on
to reset generator.
E033
GEN BATTERY LOW
Warning Lithium Battery
Voltage Low
Generator detects lithium battery
voltage is low.
Replace lithium battery.
E034
+12VDC ERROR
+12 V DC Error
+12VDC rail is out of tolerance.
Check +12VDC rail.
E035
-12VDC ERROR
-12 V DC Error
-12VDC rail is out of tolerance.
Check -12VDC rail.
E038
CAL DATA ERROR
Calibration Data Corrupt
Error
Generator detects corrupt calibration
data.
Re-calibrate X-ray tube(s).
E039
AEC DATA ERROR
AEC data corrupt
Generator detects corrupt AEC data.
Reprogram AEC data or set
factory defaults.
E041
REC DATA ERROR
Receptor data corrupt
Generator detects corrupt receptor
data.
Reprogram receptor data or
set factory defaults.
E042
TUBE DATA ERR.
Tube data corrupt
Generator detects corrupt tube data.
Reprogram tube data or set
factory defaults.