2
Installation
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 2-4
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-02 Rev. C
2.5.2
Control Console (Cont)
4.
Locate and then remove JW1 and JW2 from the console board. Refer to the appropriate figure in
chapter 1, in the section
GENERATOR LAYOUT AND MAJOR COMPONENTS.
5.
Do not discard the jumpers that were removed in the previous step. These will need to be
reinstalled if the console PREP and EXPOSE buttons must be enabled in the future.
6.
Reinstall the console base by reversing the previous steps.
Touchscreen Console
The touchscreen base is shipped as a separate item, and must be attached to the touchscreen console
before continuing with the installation. A Phillips screwdriver is required to attach the base to the
touchscreen console. A hex key has been included to adjust the tension on the tilt-arm mechanism.
1.
Carefully unpack the touchscreen console and the desk-mount base components and set the
packaging aside. Verify that all components are undamaged.
2.
Place the touchscreen console facedown on a FLAT, CLEAN, NON-ABRASIVE surface.
3.
Note the two sets of VESA (Video Equipment Standards Association) mounting holes on the back
plate and on the touchscreen console. One set of holes is spaced at 100 mm, and the other set is
spaced at 75 mm. CPI recommends the use of the 100 mm spaced VESA holes. Line up the back
plate with the touchscreen (observe orientation). Attach the base using the screws provided. See
figure 2-1.
75
m
m
(2.
95
")
mm(3.94")
100
mm(2.95")
75
1
00
m
m
(3.
94"
)
Figure 2-1: Touchscreen base back plate showing VESA mounting holes