Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-53
22.
Select
LEFT FIELD
. Toggle the button to
select
YES
or
NO
.
Click the left field on the graphic under
Fields
to select / deselect that field. Black
indicates that the field is not selected;
green indicates that the field is selected.
Click the left field on the graphic under
Fields
to select / deselect that field.
Black indicates that the field is not
selected; green indicates that the field is
selected.
3C.7.3 Receptor Setup (Cont)
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
23.
Repeat the previous step for the
CENTER
and
RIGHT
fields.
Repeat the previous step for the center
and right fields.
Repeat the previous step for the center
and right fields.
24.
Press
>>
.
25.
Select
KV, MA, MS,
and
DENSITY.
Use
the
+
or
–
buttons to select the default kV,
mA, ms, and density, respectively.
Select the default kV, mA, ms, and
density via the
Voltage (kV), Current
(mA), Time (ms),
and
Density
dialog
boxes, respectively.
Select the default kV, mA, ms, and
density via the
Voltage (kV), Current
(mA), Time (ms),
and
Density (s)
dialog
boxes, respectively.
26.
Press
<<
and
EXIT
as required to the
GEN CONFIGURATION
menu.
27.
Repeat steps 4 to 25 for the remaining
receptors. It is necessary to return to the
GEN CONFIGURATION
menu after
programming each receptor to ensure
that the updated parameters are saved to
memory.
3C.7.4 I/O Configuration
The
I/O CONFIGURATION
function allows programming the states of the exposure for the inputs and the outputs on the room interface
board.
The
I/O CONFIGURATION
menus for the membrane console are shown below.
* INPUTS [sym] *
TUBE THERMAL SW. _ _ _ _ STANDBY STATE
DOOR ITLK:
_ _ _ _
EXIT >>