CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-63
3C.7.5 AEC Setup (Cont)
Note the following regarding GenWare
®
figure 3C-18:
•
The
AEC Channel
display near the top left side of the
AEC Setup & Calibration
window shows the selected AEC channel.
•
The
Progr. Receptors
display to the right of the AEC channel display shows which receptors are programmed for the
selected AEC channel.
Definitions of
AEC SETUP
menu items applicable to the initial AEC setup.
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
CHANNEL Channel
Number
Selects the AEC channel to be programmed.
LEFT FIELD
CENTER FIELD
RIGHT FIELD
A graphic
depicting the AEC
chamber fields.
Enables or disables the ability to select the left, center, or right AEC fields.
Membrane console:
YES:
The selected field is enabled.
NO:
The selected field is disabled.
GenWare
®
:
Refer to the graphic under
Fields:
A black field select rectangle = disabled, a green field select rectangle = enabled.