1.3.4
Locating and Mounting the Generator
The main generator cabinet is self-standing and does not need to be supported. However, the installation
should meet the following requirements:
•
The floor must be flat and level.
•
The generator installation area must be clean and free of dirt or debris.
•
The installer must supply generator hold-down brackets, if required. Alternately, mounting holes have
been provided in the base of the generator. The generator may then be anchored to the floor via
these holes.
•
Sufficient room must be provided to allow access to the rear and side of the generator for installation
and service. See figure 1-1 for recommended clearances.
•
A cable trough, conduit, or raceway (1 in; 2.5 mm, diameter) should be provided from the control
console to the main cabinet to allow routing of the control cable if required.
•
The control console is normally freestanding on a desk or shelf. It may be anchored if necessary.
Figure 1-1: Generator clearances