CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-57
3C.7.4 I/O Configuration (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
BUCKY 1 RDY
Bucky 1 Ready
Programs the Bucky 1 input at J2-5 and J2-6 on the H.V. auxiliary board.
BUCKY 2 RDY
Bucky 2 Ready
Programs the Bucky 2 input at J4-5 and J4-6 on the H.V. auxiliary board.
BKY 1 START
Bucky 1 Start
Programs the Bucky 1 “start” output at J2-7 on the H.V. auxiliary board.
BKY 2 START
Bucky 2 Start
Programs the Bucky 2 “start” output at J4-1 on the H.V. auxiliary board. This is normally
programmed as the tomo output.
ROOM LIGHT
Room Light
Programs the room light output at J11-3 and J11-4 on the H.V. auxiliary board.
The inputs and outputs defined in the previous table are programmable as follows:
•
Inputs may be programmed such that the selected input is active or inactive during various states of the generator. Inactive inputs are
ignored; unused inputs should normally be programmed to be inactive.
•
Outputs may be programmed such that the relay connected to the selected output is energized or de-energized during various states of
the generator. Unused outputs should normally be programmed to be de-energized.
•
The inputs and outputs must be programmed separately for each receptor. Each receptor may have its own unique programming.
MEMBRANE CONSOLE
The
STATE
button on the upper right hand side of the menu selects the current state. The word
STATE
is preceded by the description of the
state: for example,
STANDBY
.
The arrow in the lower middle area points to the current level for the selected state. Moving to the next state is accomplished by
pressing the
STATE
button. The states are as follows:
•
STANDBY
Sets the state of the I/O when the generator is in standby or idle mode.
•
PREP
Sets the state of the I/O when the generator enters PREP mode.
•
GEN RDY
Sets the state of the I/O when the generator has completed PREP mode and is ready to expose.
•
RAD EXP
Sets the state of the I/O when the generator starts a radiographic exposure.