3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-68
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.8.0 DAP PRINTER SETUP
1.
Connect the DAP printer to the
DATA LINK
connector J4 on the rear of the console. Refer to the figure “
Rear of control console
” in
chapter 2 of the CMP 200
®
service manual for the connector location.
2.
Follow the procedure in 3C.10.2 to set up and test the DAP printer.
NOTE:
THE GENERATOR MUST BE CONFIGURED FOR THE SPECIFIC PRINTER (SEIKO INSTRUMENTS DPU-414 OR SLP-200
PRINTER). THE GENERATOR WILL BE COMPATIBLE ONLY WITH THE SELECTED PRINTER, THEREFORE ONLY THAT
PRINTER MODEL MUST BE USED IN THIS INSTALLATION.
NOTE:
THE PAPER OR LABELS USED IN THE PRINTER MUST MEET ALL APPLICABLE REGULATIONS. MEDICAL GRADE PAPER
OR LABELS, APPROVED FOR MEDICAL RECORDS, MUST NORMALLY BE USED.
3C.9.0 DAP INTERFACING
3C.9.1 DAP Compatibility
The CMP 200
®
series generator, when equipped with the DAP option, is compatible with the DAP devices listed in section 3C.10.2. The
correct DAP device must be selected in the DAP SETUP menu as described in 3C.10.2 in order to ensure device compatibility.
The DAP chamber, when fitted with the proper interconnect cable, plugs directly into the generator control board in the generator. When
ordering the DAP chamber from the DAP manufacturer, specify the CPI compatible interconnect cable, if available. This is a special cable
terminated with a 9-pin male “D” connector that is designed to plug directly into the generator control board. If this cable is not available from
the DAP device manufacturer, consult CPI product support for the required cable to connect the DAP chamber to the generator. Refer to the
table below for the CPI cable assembly part numbers.
DAP DEVICE
INTERCONNECT CABLE
PTW PX-T11020
736145-00
Gammex-RMI 841S
736146-00
VacuTEC VacuDAP 2004
Contact VacuTEC for this cable assembly.
Scanditronix 120-131
736148-00