3C.10.2 DAP SETUP Menu Items (Cont)
Use these steps to set up the DAP device. DAP calibration is done after DAP setup
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
1.
From the
GEN CONFIGURATION
menu,
select
DAP SETUP
.
From the GenWare
®
GENERATOR
UTILITIES application, select
DAP Setup
from the
Setup
menu, or use the DAP
setup button
on the GenWare
®
toolbar.
Press the
button on the
GenWare
®
toolbar to access the
DAP
Setup
utility.
2.
From
DAP SETUP
menu 1, select
DAP
.
Toggle the button to select
ON
or
OFF
.
The DAP device must be enabled if you
wish to continue.
Check the
DAP
checkbox to enable the
DAP function.
The DAP device must be enabled if you
wish to continue.
Check the
DAP
checkbox to enable the
DAP function.
The DAP device must be enabled if you
wish to continue.
3.
Select
DEVICE TYPE 1
. Use the
+
or
–
buttons to select the desired DAP 1
device.
Select the desired DAP 1 device
via the
Device Type 1
dialog box.
Select the desired DAP 1 device via the
DAP Device Type 1
dialog box.
4.
Select
TEST VALUE 1
. Use the
+
or
–
buttons to enter the test value for DAP
device 1.
Select the test value for DAP device 1 via
the
Test Value 1
dialog box.
Select the test value for DAP device 1 via
the
Test Value 1
dialog box.
5.
Select
DAP PRINTER
. Toggle the button
to select
OFF, DPU414,
or
SLP200
.
Select the desired printer
via the
DAP
Printer
dialog box.
Select the desired printer
via the
DAP
Printer
dialog box.
6.
Press
DAP
to access the DAP
RESET
,
TEST
, and
functions.
When DAP is enabled, a DAP display
window is opened immediately below the
GenWare
®
toolbar, as shown in figure
3C-19.
When DAP is enabled a DAP display
window is opened along the bottom edge
of the screen.
7.
Press
RESET
to reset the DAP display to
zero.
Press
Reset
to reset the DAP display to
zero.
Press
Reset
to reset the DAP display to
zero.
8.
Press
TEST
to test the DAP system. This
will test the DAP chamber and DAP
circuits.
Press
Test
to test the DAP system. This
will test the DAP chamber and DAP
circuits.
Press
Test
to test the DAP system. This
will test the DAP chamber and DAP
circuits.
9.
Press
to print a DAP label.
10.
Press
EXIT