Page 3C-26
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.7.1 Tube Selection / Tube Setup (Cont)
The following menu is only available if additional tubes have been downloaded via a computer running GenWare
®
.
TUBE SELECTION screen 4
* TUBE 1 SELECTION *
IF ADDITIONAL TUBES HAVE BEEN DOWNLOADED FROM GenWare
®
THEY WILL BE DISPLAYED ON THIS SCREEN
<<
RETURN
The next three menus show the default tube limits. These menus appear after a
tube has been selected in the previous steps.
TUBE DEFAULTS screen 1
*TUBE 1: RAD60 0.6/1.2 12
º
REV 1.4*
TUBE SPEED: LOW
MAX SF KW LS: 21.3
MAX LF KW LS: 58.9
+
MAX KV: 125
-
EXIT >>