Error Messages (Cont)
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touch screen console)
PROBLEM
ACTION
E103
CAL-MAN. TERM.
Calibration error –
manually terminated
Operator released exposure button
during auto calibration.
Retry auto Calibration.
E104
CAL-NO MA
Calibration error – no mA
No mA feedback detected during
auto calibration.
Call product support.
E105
CAL-MIN MA ERR.
Calibration error –
minimum mA not
calibrated
Minimum generator mA was
exceeded at start of calibration.
Reduce filament standby
current on primary and/or
secondary filament.
E150
TABLE COMM ERR
Table communication
error
A communication error has occurred
with the table.
1. Ensure that the table is
powered on.
2. Check the serial
communication cable,
connectors, etc.
E152
Table error
The table has sent a table error
message to the generator.
Refer to the table
manufacturers documentation.
E154
Table emergency stop
The table emergency stop has been
activated.
Reset the table emergency
stop.
E240
OUT OF LABELS
The printer is out of labels.
Load more labels.
E241 LABEL
JAMMED
The labels are jammed.
Clear the paper jam.
E242 PRINTER
ERROR
The printer self-diagnostics have
reported a printer problem.
Refer to the printer manual.
E243
PRINTER COMM ERR
The printer has reported a
communication error.
Try printing again.
E244 PLATEN
OPEN
The paper platen is not positioned
properly.
Check the platen position.
E245 PRINTER
OFF-LINE
The printer is off-line.
Put the printer on-line.