CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-103
3C.13.6 Touchscreen Setup (Cont)
Step Action
2.
Check the applicable item to enable that function.
3.
Press
OK
to continue or
CANCEL
to return to the system utilities menu without making any changes. If
OK
was pressed, a pop-
up window will display asking if you wish to save the changes. Press
YES
to save the changes. Press
NO
to return to the system
utilities menu;
CANCEL
cancels the changes.
Passwords
Use these steps to change the current operator and service passwords.
Step Action
1.
From the touchscreen setup window, select
PASSWORDS
.
2.
To change the operator password, press the
…
button to the right of the top operator password line. A password window will pop
up. Enter the new password.
•
Press
Clear
to cancel an incorrect entry.
•
Press
Cancel
to close the password pop-up window.
•
Press
Accept
to accept the new password.
3.
Repeat the previous step to enter the new password on the second operator password line.