CPI Canada Inc.
Installation
2
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-02 Rev. C
Page 2-30
2.8.3
Dual Speed Starter Tube Select Table (Cont)
TABLE 2-2: TUBE TYPES (DUAL-SPEED STARTER)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
CODE
Switches
1........5
H.S.
START
VOLTS
H.S.
RUN
VOLTS
H.S.
BRAKE
VOLTS
H.S.
BRAKE
TIME
BOOST
TIME
L.S.
START
VOLTS
L.S.
RUN
VOLTS
H.S.
SHIFT
CAPAC
L.S.
SHIFT
CAPAC
DUAL SPD
STARTER
PART NO.
Varian
Saphire
Std “R” stator
(20/50
Ω
)
RAD21
RAD56 0.6/1.0
RAD56 0.6/1.2
RAD60
RAD92
RAD94
10100
400
100
100
3.0 sec
2.3 sec
240
50
6 uF
37.5 uF
(-12)
31 uF
(-13)
901297-12
901297-13
901298-12
901298-13
Varian DX62
300-400 kHu,
“STD” stator
(23/56
Ω
)
A192B
A196
A197
A256
A272
A282
A286
A292
00000
400
100
100
3.0 sec
1.4 sec
240
50
6 uF
37.5 uF
(-12)
31 uF
(-13)
901297-12
901297-13
901298-12
901298-13
Varian DX62U
Universal
300-400 kHu,
configured as
“STD” or “R”
stator
(15/36
Ω
)
A192B
A196
A197
A256
A272
A282
A286
A292
00000
400
100
100
3.0 sec
1.4 sec
240
50
6 uF
37.5 uF
(-12)
31 uF
(-13)
901297-12
901297-13
901298-12
901298-13
NOTE
Two settings are shown for the Dunlee PX1400 tube with “S” stator. The settings with the 4.5 sec boost times are per older Dunlee data sheets. The
values with the 1.9 sec boost times are per newer Dunlee data sheets (May 2002). The installer must verify tube compatibility per the appropriate
Dunlee tube data sheet.