3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-40
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.7.2 Generator Limits (Cont)
Definitions of
GENERATOR LIMITS
menu items.
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
MAX KW
Max. Generator
Power (kW)
Sets the maximum generator kW limit.
MAX MA
Max. Generator
Current (mA)
Sets the maximum generator mA limit.
MIN MA
Min. Generator
Current (mA)
Sets the minimum generator mA limit.
MAX MAS
Max. Generator
Current-Time
(mAs)
Sets the maximum generator mAs limit.
ROTOR BOOST
This function is
not available in
GenWare
®
Sets the low speed starter boost time, in milliseconds. The range is 1000 ms (1 sec) to 4000 ms
(4 sec), adjustable in 100 millisecond increments.
BEFORE MAKING ANY CHANGES IN THIS SECTION, CONSULT THE X-RAY TUBE DATA SHEETS TO ENSURE THAT THE
PROPOSED CHANGES DO NOT EXCEED THE MANUFACTURERS RECOMMENDED LIMITS. THE ORIGINAL GENERATOR LIMITS
SHOULD BE RECORDED IN A COPY OF TABLE 3C-3 BEFORE CONTINUING.