CPI Canada Inc.
Troubleshooting
5
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-4
5.2.2
Limit Messages (Cont)
MESSAGE
(membrane console)
MESSAGE
(touchscreen console)
PROBLEM
ACTION
GEN DUTY WARNING
Generator duty cycle
limit
The X-ray generator has reached its duty cycle
warning limit.
Re-evaluate technique factors.
Allow generator to cool if
possible. If exposures are
continued, serious generator
damage may result due to
overheating.
GEN. JOULE LIMIT
Generator Joule Limit
The requested exposure will exceed the
generators kilo joule heat limit.
Allow the generator to cool
sufficiently to allow the
requested exposure.
GEN. KV LIMIT
Generator kV limit
Requested kV not allowed as generator kV limit
has been reached.
None.
GEN. KW LIMIT
Generator kW limit
Requested parameter not allowed as generator
kW limit has been reached.
None.
GEN. MA LIMIT
Generator mA limit
Requested mA not allowed as generator mA
limit has been reached.
None.
GEN. MAS LIMIT
Generator mAs limit
Requested mAs not allowed as generator mAs
li130
mit has been reached.
None.
GEN. MS LIMIT
Generator ms limit
Requested ms not allowed as generator ms limit
has been reached.
None.
INVALID PARAM.
Invalid communication
parameter.
Generator detected invalid parameter within
received message, message ignored.
Select valid parameter.
Parameter limit
The selected parameter has exceeded its limits.
None.
TUBE KV LIMIT
Tube kV limit
Requested kV not allowed as tube kV limit has
been reached.
None.
TUBE KW LIMIT
Tube kW limit
Requested parameter not allowed as tube kW
limit has been reached.
None.
TUBE MA LIMIT
Tube mA limit
Requested mA not allowed as tube mA limit has
been reached.
None.
TUBE MAS LIMIT
Tube mAs limit
Requested mAs not allowed as tube mAs limit
has been reached.
None.