8
Spares
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 8-6
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-10 Rev. C
8.2.0
SPARE PARTS LIST (Cont)
NOTE:
1.
Refer to table 8-1 for the part number for the H.V. auxiliary board. Four different part numbers are
used, depending on the application:
TABLE 8-1
STATOR TYPE
208 / 230 V GENERATORS
400 / 480 V GENERATORS
“R” type stator
739445-00
739442-00
GE 23/23
Ω
stator
739445-01 739442-01
Units with the dual speed starter option will be fitted with H.V. auxiliary board 739442-00.
Before installing a replacement H.V. auxiliary board, jumpers must be properly installed on this board
in order to select the Bucky voltage (24 VDC, 110 VAC, or 220 VAC). After determining the Bucky
drive voltage requirement, connect the appropriate pair of jumpers in accordance with table 8-2
*
.
The jumpers that select the Bucky voltage must be set per table 8-2.
TABLE 8.2
BUCKY
OUTPUT
H.V. AUXILIARY BOARD JUMPERS
(units with isolated 110 / 220 VAC, using auxiliary transformer 739446)
24 VDC
E11-E5, E7-E8
110 VAC
E9-E5, E7-E8
220 VAC
E12-E5, E7-E8
After determining the required rotor boost voltage, connect a wire jumper between the desired tabs on
the boards
*
.
TABLE 8-3
ROTOR BOOST
VOLTAGE
H.V. AUXILIARY BOARD JUMPERS
120 VAC
E14-E15
240 VAC
E16-E15
*
A complete set of jumpers is shipped with each new generator. Three of the jumpers are installed on
the H.V. auxiliary board, and the remaining jumpers are in a bag attached to the lip on the inside of
the cable access slot above the HT tank.
If the Bucky or low-speed starter boost voltage needs to be changed on an existing H.V. auxiliary
board, use the existing jumper(s) if they are of the correct length and have the proper connector to fit
the tab(s) on the board. Otherwise, select the shortest wires from the spare jumper set that will
connect between the desired tabs on the board and that have the proper connectors on the jumpers.
For spares boards, the jumpers will need to be removed from the “old” board and reused.
Additionally, JW1 must be set per table 8-3 before installing a replacement H.V. auxiliary board.
TABLE 8-4
UNIT TYPE
JW1 CONFIGURATION: H.V. AUXILIARY BOARD
1 phase
OPEN (jumper pins 2-3)
3 phase
CLOSED (jumper pins 1-2)