CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-67
3C.7.7 Tube Calibration (Cont)
Use these steps to perform the tube auto calibration.
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
1.
From the
GENERATOR SETUP
menu
select
GEN CONFIGURATION
.
Select
Auto Tube Calibration
from the
Setup
menu, or use the auto calibration
button
on the GenWare
®
toolbar.
Press the
button on the GenWare
toolbar to access the
Tube Calibration
utility.
2.
Select
TUBE CALIBRATION
.
3.
Select
FOCAL SPOT
. Toggle the button
to select the desired focal spot to
calibrate (
SMALL
or
LARGE)
.
Start with
SMALL
.
Under
Focus,
select the desired focal
spot to calibrate (
small
or
large)
.
Start with
small
.
Under
Focus,
select the desired focal
spot to calibrate (
small
or
large)
.
Start with
small
.
4.
Press and hold the X-RAY button (or use
the optional hand switch) to begin the
calibration procedure.
Press and hold the X-RAY button (or use
the optional hand switch) to begin the
calibration procedure.
Press and hold the X-RAY button (or use
the optional hand switch) to begin the
calibration procedure.
5.
When finished calibrating the small focus,
press
RETURN
and then repeat
calibration on the large focal spot.
Repeat the calibration on the large focal
spot.
Repeat the calibration on the large focal
spot.
6.
When auto-calibration is completed, press
EXIT
to return to the
GENERATOR
SETUP
menu.
7.
Press
EXIT SETUP
to exit out of the
setup and calibration mode and return to
the normal operation mode.
NOTE:
SHOULD AN ERROR OCCUR DURING AUTO CALIBRATION, AN ERROR MESSAGE WILL BE DISPLAYED. THE GENERATOR
WILL THEN LIMIT THE TUBES OPERATION TO THE RANGE IN WHICH IT WAS CALIBRATED, THUS ALLOWING FOR
PARTIAL OPERATION OF THE GENERATOR.