CPI Canada Inc.
Troubleshooting
5
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-9
5.2.3
Error Messages (Cont)
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touchscreen console)
PROBLEM
ACTION
E055
NO FIELDS ACTIVE
No fields selected in
AEC mode
AEC enabled but no fields are
selected.
Select AEC field(s).
E056
NO TUBE SELECTED
Receptor Disabled
All Receptors have no X-ray tube
programmed.
Program receptor(s) with tube
number.
E057
AEC STOP ERROR
AEC stop signal In
wrong state
AEC stop signal (P.T. stop signal) is
active low indicating exposure is
finished during prep state.
1. Check that the P.T. ramp
does not exceed the P.T.
reference during prep state.
2. Check AEC device for
proper operation.
E058
CONSOLE BUT ERR.
Console back-up timer
Console has detected exposure
exceeded backup time and
terminated exposure.
Call product support.
E060
EXP. KV HIGH
High kV error
kV exceeds high kV tolerance level.
1. Check the output of the kV
reference DAC on the
generator CPU board.
2. Measure the output of the
generator with a dynalyzer
or a non-invasive kVp
meter.
E061
EXP. KV LOW
Low kV error
kV exceeds low kV tolerance level.
1. Check the output of the kV
reference DAC on the
generator CPU board.
2. Measure the output of the
generator with a dynalyzer
or a non-invasive kVp
meter.
E063
FACTORY DEFAULTS
Factory defaults set
S3-1 on the generator control board
is set to the LOAD DEFAULTS
position.
Set S3-1 to the NORMAL
position. The generator will not
exit the initialization phase
until this is done.