3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-64
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.7.5 AEC Setup (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
CHAMBER TYPE
Chamber Type
Selects the AEC chamber type.
Membrane console:
ION:
Ion chamber.
S/S:
Solid-state chamber.
APL:
Apelem.
GenWare
®
:
Ion:
Ion chamber.
Solid
State:
Solid-state chamber.
Apelem:
Apelem.
FILM SCREEN 1
FILM SCREEN 2
FILM SCREEN 3
Film Screen
Active
Enables or disables the ability to select film screen 1, film screen 2, or film screen 3.
Membrane console:
YES:
The selected film screen is enabled.
NO:
The selected film screen is disabled.
GenWare
®
:
Checking
1
,
2
, or
3
under
Film Screen Active
enables that film screen.
R FIELD COMP
C FIELD COMP
L FIELD COMP
R. Field Comp
C. Field Comp
L. Field Comp
Allows left, center, and right field balance. This applies to solid-state AEC chambers only. The
calibration procedure is described in the section
AEC CALIBRATION (TABLE BUCKY
in
chapter 3D.