3D
AEC Calibration
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3D-38
Rev. A CMP 200
®
Service Manual Ch # 901476-05
3D.9.1 AEC Setup Worksheet (Cont)
Note: The example below is supplied for reference only. It does not represent an actual installation.
FUNCTION
RECEPTOR 1
RECEPTOR 2
RECEPTOR 3
Film/Screen
1.
Lanex/reg Lanex/reg Lanex/reg
2.
Lanex/med
Lanex/chest
3.
Nominal optical density:
1.2
1.1
1.4
Grid ratio/SID:
12:1
8:1
10:1
Min - max kVp range:
60 - 120
65 - 140
80 - 110
±
Density steps
±
8
±
8
±
8
Density mAs
(dose) change
+%: 12.5 12.5 12.5
(per step)
-%: 6.25 6.25 6.25
Chamber type:
Ion
Solid state
Ion
Regulatory AEC dose
requirements?
Yes
Yes Yes
Is film processing maintained?
Yes
Yes
Yes
Assigned receptor name:
Table
Wall
Aux
Are all cassettes similar?
Yes
Yes
Yes
Additional
notes:
Additional
notes:
Table 3D-12: Sample AEC setup worksheet