1
Pre-installation
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-22
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-01 Rev. C
1.3.5
Dimensions, Cable Entrance and Seismic Center Location
Figure 1-2 shows the dimensions of the generator cabinet, the locations of the cable access slots, the AC
mains cable entry, and the seismic center location for the CMP 200
®
/ CMP 200
®
DR X-ray generator.
The dimensions and weight of the generator and control console(s) are shown in the table below.
ITEM LENGTH
WIDTH
HEIGHT
WEIGHT
Main cabinet in shipping pack
30.5 (775)*
21.5 (546)*
38 (965)*
151 (68.6)
Main cabinet unpacked
25.7 (653)*
13.7 (348)*
24.3 (617)*
135 (61)
Membrane console
12.3 (313)**
10.9 (277)**
3.7 (94)**
6 (2.72)
Touchscreen control console
with base
15.9 (404)
maximum**
14.6 (371)**
11.2 (285)**
18 (8.2)
Touchscreen control console
without base
13.1 (333)**
14.6 (371)**
3.5 (89)**
10 (4.5)
* Refer to Figure 1-2
** Refer to Figure 1-3
The above dimensions are inches (mm); weights are in pounds (kg).
Figure 1-2: CMP 200
®
/ CMP 200
®
DR cable entry locations and seismic center