Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-20
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.5.4 Console (Cont)
Definition of console parameters as used in this section.
FUNCTION
(MEMBRANE
CONSOLE)
DESCRIPTION
SLOW KEY
REPEAT
Determines the speed at which displays change while the selected key is pressed for the first 5 counts.
MED. KEY
REPEAT
Determines the speed at which displays change while the selected key is pressed for the next 5 counts.
FAST KEY
REPEAT
Determines the speed at which displays change while the selected key is pressed after 10 counts.
SPEAKER
VOLUME
Sets the speaker volume for the control console in the range
1
to
15.
LCD SCREEN
Toggles between normal and reverse video for the LCD display.
APR MODE
NO
allows the operator to select an APR view, and still have the ability to manually select receptors, focus, technique,
film screen, AEC fields, etc.
YES
allows the operator to select all of the above
EXCEPT
the technique selection (AEC, mAs, mA/ms) i.e. this
disables the ability to select AEC, mAs, mA/ms in APR mode.
AEC, mAs, mA/ms changes can only be made by selecting an APR technique that has been programmed to the
desired technique.
LOAD CONSOLE
DEFAULTS?
YES:
Initializes the console CPU’s NVRAM to the factory default settings when the generator is switched ON. This
restores the factory defaults for the APR and the CONSOLE settings.
NO:
The NVRAM is not reset when the generator is switched on.
The normal setting for this function is NO. Do not set to YES unless you intend to restore the console factory
defaults. Doing so will cause all custom console and APR settings to be lost.
LOGO ON?
YES:
The predefined logo is displayed briefly after the generator is switched on.
NO:
The logo is not displayed.
LANGUAGE
Selects the language for status and error messages (the APR text must be changed via the CPI GenWare
®
utility
software).