3C.6.0 APR EDITOR (Cont)
Definition of the
APR EDITOR
function.
FUNCTION
(MEMBRANE
CONSOLE)
DESCRIPTION
APR EDITOR
Enables / disables the ability of the operator to make
and then save
changes to APR techniques.
ENABLED:
Allows the operator to change the default APR technique(s), and then save the changes to
memory. The APR will subsequently default to the changed technique.
DISABLED:
Allows temporary editing of APR technique(s), but does not allow the changes to be saved to
memory. The APR will always default to the original technique when the generator is switched
OFF and then ON again.
The generator stores the last APR EDITOR setting before being switched off. If the APR editor was
previously ENABLED, APR changes may subsequently be made and then saved in normal operating
mode without the need to manually set the APR editor to ENABLED. To disable APR technique
changes, the APR editor must be set to DISABLED.
NOTE:
APR text may be altered by using a computer running GenWare
®
. Further documentation regarding this function is included
with GenWare
®
in the form of an MS Word document.
This function does not apply to GenWare
®
.
Use these steps to set the
APR EDITOR
. Refer to the definition in the previous table.
Step
Action (membrane console)
1.
From the
GENERATOR SETUP
menu, select
APR EDITOR
.
2.
Toggle the button to select
ENABLED
or
DISABLED
.