CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-47
3C.7.3 Receptor Setup (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
RECEPTOR SYM
This is a console
function only; does
not apply to
Genware.
Allows one of the predefined receptor symbols
[sym]
to be assigned to the selected
receptor.
MEMORY Memory
Defines the techniques that will be defaulted to when a receptor is selected.
YES / on:
The selected receptor will remember it’s last techniques such that those
techniques are displayed when that receptor is re-selected.
NO / off:
The selected receptor will not remember the last techniques used on that
receptor. The techniques used will be the same as last used on the
previous receptor.
DEF /
default:
The techniques used for that receptor will be as programmed. See receptor
setup menus 3 and 4 (membrane console) or the
Receptor Defaults
tab
(GenWare
®
).
TOMO Tomo
Enables or disables tomographic operation.
TOMO BUT
Tomo Back-Up Time
(ms)
Sets the tomo backup time.
SF/LF SWITCH
This function is not
available in
GenWare
®
Enables or disables the ability of the generator to automatically select the large or small
focus.
AUTO:
Small or large focus will automatically be selected by the generator. The
small to large (and vice-versa) switching will occur at the
MAX SF MA
set
point.
MAN:
The operator must manually select small or large focus.