CPI Canada Inc.
Troubleshooting
5
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-07 Rev. B
Page 5-11
5.2.3
Error Messages (Cont)
ERROR
CODE
MESSAGE
(membrane console)
MESSAGE
(touch screen console)
PROBLEM
ACTION
E082
KV OVER VOLTAGE
kV Over Voltage
The output kV exceeded the
maximum allowed kV limit.
1. Auto calibrate tube (per
TUBE CALIBRATION
in
chapter 3C)
*
.
2. Consult product support.
E083
ANODE MA FAULT
Anode mA Fault
The anode current exceeded the
maximum allowed limit.
1. Check for arcing of the X-
ray tube.
2. Check for arcing in the HT
tank; if failure of HT tank is
suspected, contact product
support.
E084
CATHODE MA FAULT
Cathode mA Fault
The cathode current exceeded the
maximum allowed limit.
1. Check for arcing of the X-
ray tube.
2. Check for arcing in the HT
tank; if failure of HT tank is
suspected, contact product
support.
E085
ROTOR MAIN FAULT
Rotor Main Fault
Indicates that insufficient current was
sensed in the stator main winding.
1. Check the X-ray tube stator
wiring.
2. Check the rotor board.
E086
ROTOR SHFT FAULT
Rotor Shift Fault
Indicates that insufficient current was
sensed in the stator shift winding.
1. Check the X-ray tube stator
wiring.
2. Check the rotor board.
E100
CAL-MAX MA ERR.
Calibration error –
maximum mA exceeded
Maximum mA has been exceeded
during auto calibration.
Repeat auto calibration and/or
decrease standby current.
E101
CAL-DATA LIMIT
Calibration error –
calibration data table
exceeded
Auto calibration has exceeded data
table length due to an excessive
number of exposures.
1. Check to see if the filament
standby current is too low.
2. Retry auto calibration.
E102
CAL-MAX FIL ERR
Calibration error –
maximum filament
current exceeded
Maximum filament current for the
selected focus has been reached.
1. Check to see if the
maximum filament current
limit can be increased.
2. Retry auto calibration.