1
Pre-installation
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 1-18
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-01 Rev. C
1.3.2
Generator Power Requirements (Cont)
The table below depicts the power requirements for the 80 kW CMP 200
®
DR X-ray generators.
Line Voltage
400 VAC
±
10%, 3 phase.
480 VAC
±
10%, 3 phase.
Line Frequency
50/60 Hz.
Momentary Current
155 Amps/phase at 400 VAC.
130 Amps/phase at 480 VAC.
Nominal Current
*
≤
5 Amps.
Momentary Power Consumption
105 kVA.
*
Nominal Current = Generator standby current only. External or installer-supplied equipment
connected to the generator may increase the nominal current beyond the values shown.
The following table defines the power line requirements for the generators.
NOTE: THE FOLLOWING TABLE CONTAINS RECOMMENDED VALUES FOR THE WIRE SIZES BETWEEN
THE MAINS DISCONNECT AND THE GENERATOR. THE ACTUAL VALUES USED AT AN
INSTALLATION ARE DEPENDENT ON THE QUALITY OF THE INPUT LINE (VOLTAGE LEVEL), THE
CURRENT REQUIREMENTS, AND THE LENGTH OF THE CABLE RUN, AND MUST BE CONFIRMED
BY THE INSTALLER.
FINAL SELECTION OF GENERATOR INPUT WIRE AND DISCONNECTS AS WELL AS THE
CABLING FROM THE DISTRIBUTION TRANSFORMER TO THE MAINS DISCONNECT MUST MEET
THE REQUIREMENTS OF THE LOCAL ELECTRICAL CODES, AND IS USUALLY DETERMINED BY
HOSPITAL / CONTRACTOR ENGINEERING.
THE RATINGS LISTED CONSIDER THE GENERATOR REQUIREMENTS ONLY. THE INSTALLER
MUST MAKE THE NECESSARY COMPENSATION FOR ADDITIONAL LOAD REQUIREMENTS.
A POOR QUALITY INPUT LINE MAY RESULT IN THE INSTALLER HAVING TO
DERATE THE GENERATOR'S MAXIMUM POWER.