CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-21
3C.5.4 Console (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
DESCRIPTION
CM THICKNESS
ON?
YES:
CM THICKNESS mode is enabled.
NO:
CM THICKNESS mode is disabled.
If CM THICKNESS is enabled, the AEC BACKUP mode (under RECEPTOR SETUP) must be set to
mAs
for each
image receptor. If this is not done, CM thickness will not be available in AEC mode.
PASSWORD
Allows the programming mode password to be changed.
This function does not apply to GenWare
®
.
Use these steps to set the console parameters. Refer to the definitions in the previous table.
Step
Action (membrane console)
1.
From the
UTILITY
menu select
CONSOLE.
This accesses CONSOLE CONFIG screen 1.
2.
Select
SLOW KEY REPEAT.
Use the
+
or
–
buttons to set the “slow key repeat” time.
3.
Select
MED. KEY REPEAT.
Use the
+
or
–
buttons to set the “med. key repeat” time.
4.
Select
FAST KEY REPEAT.
Use the
+
or
–
buttons to set the “fast key repeat” time.
5.
Select
SPEAKER VOLUME.
Use the
+
or
–
buttons to set the speaker loudness.
6.
Select
LCD SCREEN.
Toggle the button to select normal or reverse video.
7.
Select
APR MODE.
Toggle the button to select
YES
or
NO
.
8.
Press
>>
.
9.
Do not perform this step unless you intend to restore the console factory defaults.
To restore the console factory default settings, select
LOAD CONSOLE DEFAULTS?
and then toggle the adjacent selection
button to select
YES.
In order for the changes to take effect, the generator must be switched OFF and then ON again. The console will prompt for a
YES
or
NO
to loading defaults when it is powered on again. Select
YES
to both prompts to reset the console and APR defaults.
Selecting
NO
will not update the defaults.
The
LOAD CONSOLE DEFAULTS
setting automatically resets to
NO
the next time the generator is switched on.
10.
Select
LOGO ON?.
Toggle the button to select
YES
or
NO
.
11.
Select
LANGUAGE.
Toggle the button to select the desired language for status and error messages.