CPI Canada Inc.
Interfacing and Programming
3C
Use and disclosure is subject to the restrictions on the title page of this CPI document.
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
Page 3C-59
3C.7.4 I/O Configuration (Cont)
Use these steps for programming the I/O functions. Refer to the definitions in the previous table.
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
1.
From the
GEN CONFIGURATION
menu,
select
I/O CONFIGURATION
.
From the
Receptor Setup
window, select
the
Inputs
tab.
From the
Receptor Setup
window, select
the
Inputs
tab.
2.
Select the first receptor to be
programmed.
Select the first receptor to be
programmed. Use the
Receptor
tabs on
the
Receptor Setup
window.
Select the first receptor to be
programmed. Use the
Receptor
buttons
on the
Receptor
window.
3.
Press the
STATE
button to select the first
state that can be programmed for the
TUBE THERMAL SW.
input. This is the
STANDBY
state.
Toggle the
TUBE THERMAL SW.
button
to select the desired logic level (low or
high) to disable or enable that input
during the standby state.
For the
Tube Thermal Switch
input,
identify the first state that can be
programmed. This is the
Standby
state.
Enable or disable the
Tube Thermal
Switch
input during the standby state by
checking (with an
X
), or unchecking the
Standby
column to the right of
Tube
Thermal Switch.
For the
Tube Thermal Switch
input,
identify the first state that can be
programmed. This is the
Standby
state.
Enable or disable the
Tube Thermal
Switch
input during the standby state by
checking (with an
X
), or unchecking the
Standby
column to the right of
Tube
Thermal Switch.
4.
Repeat the previous step for each state.
Repeat the previous step for each state.
Repeat the previous step for each state.
5.
Repeat steps 3 and 4 for each input. Use
the
>>
button to advance to
INPUTS
menu 2.
Repeat steps 3 and 4 for each input.
Repeat steps 3 and 4 for each input.
6.
Press
>>
to select the
OUTPUTS
menu.
Select the
Outputs
tab.
Select the
Outputs
tab.
7.
Repeat steps 3 and 4 to program each
output.
Repeat steps 3 and 4 to program each
output.
An unchecked state causes the relay
connected to that output to be de-
energized during the selected state. A
checked state (with an
X
) will result in the
relay being energized during the selected
state.
Repeat steps 3 and 4 to program each
output.
An unchecked state causes the relay
connected to that output to be de-
energized during the selected state. A
checked state (with an
X
) will result in the
relay being energized during the selected
state.
8.
When finished the I/O programming for
the current receptor, exit to the
GEN
CONFIGURATION
menu.
Select
Apply
to save the programming for
the current receptor.
Select
Apply
to save the programming for
the current receptor.