3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-14
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.5.1 Setting Time And Date (Cont)
Step
Action (membrane console)
Action (PC GenWare
®
)
Action (Touchscreen GenWare
®
)
8.
To synchronize GenWare
®
to the clock in
your computer, check the
Auto
box in the
lower left corner of the
Date and Time
window.
Doing so will transfer the computers time
and date settings to the time and date
dialog boxes in the upper half of the
Date
and Time
window. Clicking on
Apply
will
apply those settings to the generator’s
clock.
To synchronize GenWare
®
to the clock in
your touchscreen, press the
Synchronize
button in the lower left corner of the
Date
& Time Control Window
.
Doing so will transfer the Touchscreen
computer’s time and date settings to the
time and date dialog boxes in the upper
half of the
Date and Time Control
Window
. Pressing
Apply
will apply
those settings to the generator’s clock.
NOTE: The time does not increment when in the SET TIME & DATE mode.
9.
Press
EXIT
to return to the
UTILITY
menu.
Press
Close
to exit the
Date and Time
window.
Press
Close
to exit the
Date & Time
Control Window
.
3C.5.2 Error Log
This utility allows display of the error messages stored in the generator’s error log. The error log storage limit is 200 messages at which point
the generator will replace the oldest with the most recent.
The
ERROR LOG
menu for the membrane console is shown below.
* ERROR LOG *
ERROR # 1 OF 18
DATE: 8-19-2002
TIME: 13:09
ERROR CODE: 200
+
ERROR MESSAGE: ANODE HEAT WARN -
KVp:
80
mA
:
100
ms:
200
EXIT