3C
Interfacing and Programming
CPI Canada Inc.
Use and disclosure is subject to the restrictions on the title page of this CPI document.
Page 3C-48
CMP 200
®
X-Ray Generator Service Manual Ch # 901476-04 Rev. C
3C.7.3 Receptor Setup (Cont)
FUNCTION
(MEMBRANE
CONSOLE)
FUNCTION
(GenWare
®
)
DESCRIPTION
AEC BACKUP
AEC Mode
(AEC Back-up
Mode)
Defines the AEC backup mode to be used.
FIXED:
The generator will determine the maximum AEC backup time, not to
exceed preset AEC backup mAs/ms values or system limits. The
characters
AEC
will be displayed in the time window of the LED
display during AEC operation.
MAS:
Allows the operator to adjust the AEC backup mAs, not to exceed
preset AEC backup mAs/ms values or system limits. The mAs value
will be displayed in the time window of the LED display during AEC
operation.
MS:
Allows the operator to adjust the AEC backup ms, not to exceed preset
AEC backup mAs/ms values or system limits. The ms value will be
displayed in the time window of the LED display during AEC operation.
AEC BACKUP
MAS
AEC Back-Up mAs
(AEC Backup mAs)
Sets the maximum AEC backup mAs, to a limit of 600 mAs.
AEC BACKUP MS
AEC Back-Up ms
(AEC Backup ms)
Sets the maximum AEC back-up ms.
RAD HANG
Rad Hangover (s):
Sets the time that the rotor will continue to spin after a rad exposure has terminated.