Operation using HART communicator
Azbil Corporation
7-28
Model MTG11A/18A, MTG11B/18B, MTG14C
Manual Zero
This function is used to improve flow measurement more accurately when the flow rate becomes
25% or less of setting range. The MTG Flowmeter has three manual zeroing functions for each
excitation current. To execute the manual zero, proceed as follows.
1.
From online menu,
Select:
1. Device setup
2. Diag/Service
3. Calibration
3. Manual zero
Figure 7-49
is then displayed
.
Figure 7-49
2.
Once the display as shown Figure 7-49 appears, move the arrow key up or
down to select a manual zero to be adjusted.
3.
Select the manual zero, then the screen
as shown in Figure 7-50 appears.
Figure 7-50
4.
To continue the manual zero, press OK. Then the screen as shown in Figure 7-
51 appears. Press OK to continue. The notice as shown in Figure 7-52
appears.
Figure 7-51
Figure 7-52
5.
Once the display as shown
Figure 7-53
appears, confirm the value appears on the screen
and select an appropriate command by moving the arrow key up or down. After marking
a selection, press F4 (ENTER). Check the adjustment value by selecting “3 Refresh %”
command. If you finish the manual zero, select “4 Quit(Manual zero end) command and
press F4 (ENTER). Then screen as shown in
Figure 7-54
appears. Press OK and then
return to the Manual zero menu.
CM2-MTG300-2001.book 28 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...