Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
6-67
6
or
Press the
key and enter into
the CORRECT DAC screen.
Current value shown in the bottom
right screen presents the
adjustment variation.
By pressing the
key, you can
change from 0.002, 0.01, 0.05 to
0.25 mA.
Adjust by pressing the
key or
key.
7
Press the
key.
Display return back to the screen
shown to the right. # will be shown
in the bottom right screen. This
shows that it is still in constant
current generation mode. You need
to clear this mode.
8
Press the
key again.
9
Press the
key.
Confirm that the “#” mark has been
deleted.
10
Adjust to 100% (20 mA) by
following the same procedure as
for adjusting 0% (4 mA).
Step
Key
Procedure
SFC screen