Operation using HART communicator
Azbil Corporation
7-32
Model MTG11A/18A, MTG11B/18B, MTG14C
Pulse output adjustment
To adjust the pulse output, proceed as follows:
1.
From online menu,
Select:
1. Device setup
2. Diag/Service
3. Calibration
5. Pulse trim
Figure 7-62 will then be displayed.
If the control system is not affected by
the forcible change of the current
signal, press F4 (OK). By pressing F3
(ABORT), the procedure is canceled.
Figure 7-62
2.
Figure 7-63 will then appears. Connect
the device to measure the pulse output,
and then press F4 (OK).
Figure 7-63
3.
Figure 7-64 will then appear. By
pressing F4 (OK), the pulse output
adjustment will start and the converter
will output a pulse of 90 Hz. If it is
fine, press F4 (OK).
Figure 7-64
4.
Figure 7-65 will then appear. When
executing the adjustment, select
“SET”, and then press F4 (ENTER).
Figure 7-65