Azbil Corporation
Instrument installation
Model MTG11A/18A, MTG11B/18B, MTG14C
2-15
Accessory parts for installation
Introduction
The following parts are necessary for the installation of the detector:
• Centering nuts (standard accessory: 4 pcs.)
• Through bolts and nuts (option)
• Gaskets:
Gaskets are to be provided by the customer when using grounding
rings made of SUS material.
Gasket are supplied as a standard accessory, when using
grounding rings made of hastelloy, titanium, tantallum, or
platinum.
• Protective plate: Required when connecting the detector to polyvinyl chloride
(PVC) piping
Centering nuts
To install the detector, use centering nuts to ensure the exact center line alignment of
the pipe and the detector.
Slip the centering nuts onto the through-bolts, and set the detector on top of the nuts so
that the nuts are on four sides of the detector.
The positions of the centering nuts depend on the direction in which the detector is
installed.
For the positions of the centering nuts, refer to Figure 2-13 and Figure 2-14.
Figure 2-13 Horizontal centering of the detector
(Position two centering nuts against each flange)
Figure 2-14 Vertical centering of the detector
(Position the four centering nuts on the bottom flange)
Flange
Centering nut
position
Flange
Centring nuts
position
CM2-MTG300-2001.book 15 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...