Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-45
5-6-4 : Setting detector factor
Set the detector factor. The C1 value is always 1.0000. Set the C2 value to that shown
on the detector nameplate under the heading Detector Factor.
Step
Procedure
Screen
1
Enter ENGINEERING MODE (see “Entering
Engineering Mode and Maintenance Mode” on
page 5-36
). Then press the or key to display
the screen at right.
2
Press the key to set the detector factor.
C1 is always set to 1.0000.
Set the C2 factor to that stamped on the nameplate
of the flowmeter.
3
Press the key to move the cursor to the position
under #.
Press the MODE key to return to the
MEASURING MODE and to save the data.
20.0 %
_# C1 1.0000
C2 1.0234
20.0 %
_# C1 1.0000
C2 1.0234
20.0 %
_# C1 1.0000
C2 1.0234
CM2-MTG300-2001.book 45 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...