![Azbil MagneW Neo PLUS Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/azbil/magnew-neo-plus/magnew-neo-plus_user-manual_3042666068.webp)
Instrument installation
Azbil Corporation
2-14
Model MTG11A/18A, MTG11B/18B, MTG14C
~Note
Never attempt to force the detector between two piping flanges when the
space is too narrow. It can damage the detector.
~Note
Tighten each bolt a little at a time and apply uniform torque to all the
bolts. If leakage does not stop on completion of fastening, make sure that
the pipe is not off center, then tighten each bolt little by little. Install the
detector carefully and ensure continue to that the fastening torque does
not exceed the prescribed limit, otherwise the detector could be damaged.
m
CAUTION
Before installing the detector make sure that the pipe is exactly straight and
centered. Any irregularity in these respects could cause leakage or other hazards.
Figure 2-11 Examples of unacceptable installations (1)
Figure 2-12 Examples of unacceptable installations (2)
m
WARNING
Ensure the bore diameters of the pipe and the detector are exactly the same, install
the detector so that the gasket does not protrude into the internal bore of the pipe, as
this could result in leakage or other hazards.
Tilted pipe
Off-center
CM2-MTG300-2001.book 14 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...