![Azbil MagneW Neo PLUS Скачать руководство пользователя страница 133](http://html1.mh-extra.com/html/azbil/magnew-neo-plus/magnew-neo-plus_user-manual_3042666133.webp)
Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-27
Step
Procedure
Screen
9
(Con-
tin-
ued)
(Branch 4)
• When the result is (4) in Step 8
Check that the empty or scale on electrode status
is not detected when SENSITIVITY MID has
been set.
(Result)
• If the empty or scale on electrode status is not
detected, use that SENSITIVITY MID setting
without change.
• If the empty or scale on electrode status is
detected and the screen shown on the right
appears, this function is not available in that
environment. Set the function to OFF.
(Branch 5)
• When the result is (5) in Step 8
Check that the empty or scale on electrode status
is not detected when SENSITIVITY HIGH has
been set.
(Result)
• If the empty or scale on electrode status is not
detected, use that SENSITIVITY HIGH setting
without change.
• If the empty or scale on electrode status is
detected and the screen shown on the right
appears, this function is not available in that
environment. Set the function to OFF.
0.0 %
EMPTY OR SCALE
ON ELECTRODE
0.0 %
EMPTY OR SCALE
ON ELECTRODE
CM2-MTG300-2001.book 27 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...