Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
6-13
Writing on non-volatile memory
After downloading the changed setting data using SFC, save the setting data to
MangeW Two-wire PLUS non-volatile memory in approx. 30 seconds. Therefore, do
not turn the power off during the operation.
If you want to save the data immediately, press the
key and
key, then the
data will be written forcibly to non-volatile memory.
Changing communication method
Two wired magnetic flowmeter has the following 4 communication methods:
• SFN.A… SFC communication (Analog)
• SFN.D… SFC communication (Digital)
• HART… HART communication
• NONE… not using any communication function
~Note
SFC communication (Digital) indicates Enhanced DE communication of
Honeywell Co.
For SFC communication, select “SFN.A”.
To change the communication method, operate using the touch keys. From the data
setting screen “COM SELECT” of “OPERATOR'S MODE”, it is required to set the
communication method to SFN.A. (Figure 6-4)
Figure 6-4 Data setting screen
SHIFT
ENTER
( Yes )
NON-VOL
*
COM SELECT
SFN.A
CM2-MTG300-2001.book 13 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...