Azbil Corporation
Operation using the data setting device
Model MTG11A/18A, MTG11B/18B, MTG14C
5-73
~Note
•
If the main display shows -2.0%, the zero point value may exceed
-2.0%. Execute Auto zero before manual zero.
Step
Procedure
Screen
5
Adjust zero point by pushing the or key so that the
main display shows 0.0%.
By pushing the key once, READY changes to UP
and the zero point value increases 0.05%.
By pushing the key once, READY changes to
DOWN and the zero point value decreases 0.05%.
It takes about 20 seconds to change the zero point
value. During manual zeroing, the value in the main
display flashes. If you push the or key again, wait
until the value in the main display does not flash.
Note
Do not keep on pushing the or key.
Manual zeroing does not work.
6
Move the cursor to the mode indicator by
pushing key.
7
Push the key and display the MANUAL
ZERO2 screen.
8
Execute MANUAL ZERO2, as well as
MANUAL ZERO1. (Refer to the Step 2 to 6.)
9
Push the key and display the MANUAL
ZERO3 screen.
10
Execute MANUAL ZERO3, as well as
MANUAL ZERO1. (Refer to the Step 2 to 6.)
11
Push the MODE key and return to the
MEASURING MODE.
0.0 %
> MANUAL ZERO1
READY
0.0 %
> MANUAL ZERO1
UP
0.0 %
> MANUAL ZERO1
DOWN
0.0 %
> MANUAL ZERO1
READY
0.5 %
> MANUAL ZERO2
WORKING
0.0 %
> MANUAL ZERO2
WORKING
0.5 %
> MANUAL ZERO3
WORKING
0.0 %
> MANUAL ZERO3
WORKING
0.5 %
MEASURING MODE
CM2-MTG300-2001.book 73 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...