Setting/changing damping time constant: DAMP key
Introduction
Damping time constant is a response time of the primary delay (65.2% response) for a
step response of the flow rate. When the output fluctuations are large increase the
damping. A large damping value stabilizes the output but lowers the response
performance. The damping time constant can be set to 0.5 up to 199.9 sec. using
numeric keys.
Procedure
Use the following procedure to set the damping time constant.
Step
Key
Procedure
SFC screen
1
Confirm that the SFC is set to
“READY”.
If it is not, press the
key to set it
to “READY”.
2
Press the
key.
Result:
• The currently set damping time
constant appears as shown here.
3
Use numeric keys to set the damping
time constant. (Input range: 0.5 to
199.9)
Result:
• The changed setting is written into
the database of the converter and
displayed on the screen.
4
Press the
key to return to the
screen in step 1.
CLR
( No )
CLR
( No )
SFCM00006009D
C
DAMP
C
DAMP
SFCM00006056D
SFCM00006057D
SFCM00006058D
CLR
( No )
CLR
( No )
CM2-MTG300-2001.book 23 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...