Operation using the data setting device
Azbil Corporation
6-26
Model MTG11A/18A, MTG11B/18B, MTG14C
Setting output range and correction coefficient: URV key
Introduction
The output range of the converter is set at the factory according to the ordered
specifications. This setting can be displayed on the screen of the SFC or changed.
Definition
• URV (Upper Range Value) refers to a measured value of flow rate when the output
of the converter becomes 100% (20 mA DC in the case of analog output) and
means an upper range value of the output range of the converter. Pressing the
key displays the set URV (e.g., 10,000 m³/h) on the screen. URV setting
range is 0.3 to 10 m/s in flow rate conversion.
Procedure
Use the following procedure to display or change the set output range.
Step
Key
Procedure
SFC screen
1
Make sure that the SFC is set to
“READY”.
If it is not, press the
key to set it
to “READY”.
2
Press the
key.
Result:
• The set URV appears as shown to
the right.
3
Use numeric keys and
key to
enter the URV to be set.
4
Press the
key.
5
When pressing the
key,
correction coefficient will be set. Set if
necessary.
6
Press the
key.
7
Press the
key to return to step 1.
URV F
100%
CLR
( No )
CLR
( No )
SFCM00006009D
URV F
100%
URV F
100%
SFCM00006036D
Q
A
DE
Q
A
DE
SFCM00006037D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006037D
G
SET
G
SET
SFCM00006074D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
CLR
( No )
CLR
( No )
CM2-MTG300-2001.book 26 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...