Maintenance and troubleshooting
Azbil Corporation
8-2
Model MTG11A/18A, MTG11B/18B, MTG14C
8-1 : Maintenance and inspection of parts
8-1-1: Replacement of indicator / data setting device
Procedure
Replacement procedure is as follows.
Figure 8-1 Replacement of indicator/data setting device (with the cover removed)
~Note
When unpacking the detector, do not open the packing in the
location with high temperature and humidity, many dust and
corrosive gas atmosphere.
Step
Procedure
1
Power off the converter by circuit breaker etc.
2
The converter front cover is fixed by hexagon socket head screws (M3).
Loosen these screws with an Allen wrench (1.5).
3
Remove the converter front cover by turning it counterclockwise with
the dedicated tool.
~Note
Remove the front cover straight and with care.
4
Remove the 3 fixing screws.
5
Remove by pulling out.
6
Align the new card's connector to the converter connector.
~Note
Ensure the correct panel direction. The panel attaching
direction can be chosen from two options according to the
attaching position of this instrument.
7
Fix the card again with the 3 screws.
~Note
The connector is connected firmly by tightening the screws.
8
Attach the front cover.
~Note
Take care not to injure your fingers on the cover edge or the
thread in the case.
CM2-MTG300-2001.book 2 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...