Electrical wiring
Azbil Corporation
3-6
Model MTG11A/18A, MTG11B/18B, MTG14C
Grounding
The grounding is essential for flow measurement.
The most effective grounding method is direct connection to earth ground with
minimal impedance.
For the grounding terminal, carry out grounding work (grounding resistance 100
Ω
or
less) according to Figure 3-2 or Figure 3-3. Do not ground internal & external at the
same time.
<Integral type>
<Integral type>
<Remote type>
<Remote type converter> <Remote type detector>
Figure 3-2 Grounding
procedure that uses internal
grounding terminal
Figure 3-3 Grounding procedure that uses external
grounding terminal