Operation using the data setting device
Azbil Corporation
6-52
Model MTG11A/18A, MTG11B/18B, MTG14C
Setting diameter of detector: [DIAMETER=] function
Introduction
The diameter of the detector of the converter is set at the factory according to the
ordered specifications. The setting of this diameter can be changed.
Used when
When only the detector is replaced with one of a different diameter, this function is
used to reset the diameter.
Procedure
Use the following procedure to set the diameter of the detector.
Step
Key
Procedure
SFC screen
1
Confirm that the SFC is set to
“READY”.
If it is not, press the
key to set
it to “READY”
2
Press the
key to access the
CONFIG functions.
3
or
Press the
key or
key to
display the screen as shown on at the
right.
4
Press the
key.
5
or
Press the
key or
key to
display the screen as shown on the
right.
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
B
CONF
SFCM00006018D
H
NEXT
L
PREV
H
NEXT
L
PREV
SFCM00006019D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006020D
H
NEXT
L
PREV
H
NEXT
L
PREV
SFCM00006028D
CM2-MTG300-2001.book 52 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...