Operation using the data setting device
Azbil Corporation
6-46
Model MTG11A/18A, MTG11B/18B, MTG14C
Changing flow rate display: [DISP] function
Introduction
It is possible to set whether an instantaneous flow rate displayed on the display panel
of the converter should be expressed as real flow rate or percentage.
Percent display refers to a percentage (%) over the maximum flow rate set by the URV.
Procedure
Use the following procedure to set or change flow rate display.
Step
Key
Procedure
SFC screen
1
Confirm the SFC is set to “READY”.
If it is not, press the
key to set
it to “READY”.
2
Press the
key to access the
CONFIG functions.
3
Press the
key.
4
or
Press the
key or
key to
display this screen.
5
Press the [ENTER] key.
Result:
• The screen as shown on the right
appears and the set flow rate
display is confirmed.
6
Pressing the
key displays
DISP=%, DISP=FLOW RATE and
DISP=TOTAL by turns. Display the
screen that you want to set.
CLR
( No )
CLR
( No )
SFCM00006009D
B
CONF
B
CONF
SFCM00006018D
ENTER
( Yes )
NON-VOL
ENTER
( Yes )
NON-VOL
SFCM00006020D
H
NEXT
L
PREV
H
NEXT
L
PREV
SFCM00006050D
ENTER
( Yes )
NON-VOL
SFCM00006051D
MENU I
ITEM
DE CONF.
MENU I
ITEM
DE CONF.
SFCM00006052D
CM2-MTG300-2001.book 46 ページ 2015年9月29日 火曜日 午前10時14分
Содержание MagneW Neo PLUS
Страница 16: ...List of Figures Tables CM2 MTG300 2001 book 4...
Страница 92: ...MEMO CM2 MTG300 2001 book 38...
Страница 106: ...MEMO CM2 MTG300 2001 book 4...
Страница 266: ...MEMO CM2 MTG300 2001 book 82...
Страница 312: ...MEMO CM2 MTG300 2001 book 46...
Страница 336: ...MEMO CM2 MTG300 2001 book 22...
Страница 338: ...CM2 MTG300 2001 book 2...